RecallWatchUSA

Home / Recalls / Medical Devices

Inpeco S.A. recalls Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multipl

Recalled June 10, 2020 · FDA recall Z-2200-2020 · Inpeco S.A

Hazard
In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module. · FDA Class II
Units
61 systems
Sold at
US Nationwide distribution including in the states of IL, NY.

Description

In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.

Product
Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.
Manufacturer(s)
Inpeco S.A.
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Inpeco S.A. recalls Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multipl | RecallWatchUSA