RecallWatchUSA

Home / Recalls / Medical Devices

Inpeco S.A. recalls FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio

Recalled November 7, 2018 · FDA recall Z-0361-2019 · Inpeco S.A

Hazard
The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes. · FDA Class II
Units
679
Sold at
Foreign distribution to Netherlands, Italy, Belgium, Austria, France, Spain, and Germany.

Description

The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes.

Product
FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre- Analytical and Post-Analytical processing,
Manufacturer(s)
Inpeco S.A.
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Inpeco S.A. recalls FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a3600 and Aptio | RecallWatchUSA