Insulet Corporation recalls 18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Ass

Recalled November 23, 2022 · FDA recall Z-0154-2023 · Insulet

Hazard
The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM. · FDA Class I
Units
248,288 US; 167,802 OUS
Sold at
Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

Description

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Product
18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.
Manufacturer(s)
Insulet Corporation