Integra LifeSciences Corp. recalls C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Recalled December 18, 2019 · FDA recall Z-0613-2020 · Integra LifeSciences
Hazard
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying. · FDA Class II
Units
820
Sold at
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Description
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
Product
C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Manufacturer(s)
Integra LifeSciences Corp.