RecallWatchUSA

Home / Recalls / Medical Devices

Integra LifeSciences Corp. recalls Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization devic

Recalled November 20, 2013 · FDA recall Z-0210-2014 · Integra LifeSciences

Hazard
An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure. · FDA Class II
Units
37 units
Sold at
US Distribution including the states of AL, AR, CA, CT, FL, GA, IA, IL, MS, NY, OH, SC, TN, and TX.

Description

An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.

Product
Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization device, designed to provide rigid skeletal fixation. The Skull Clamp is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary
Manufacturer(s)
Integra LifeSciences Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Integra LifeSciences Corp. recalls Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization devic | RecallWatchUSA