RecallWatchUSA

Home / Recalls / Medical Devices

Integra LifeSciences Corp. recalls Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder Syst

Recalled November 19, 2014 · FDA recall Z-0197-2015 · Integra LifeSciences

Hazard
A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be. · FDA Class II
Units
12 units
Sold at
Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.

Description

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

Product
Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.
Manufacturer(s)
Integra LifeSciences Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov