Integra LifeSciences Corp. recalls K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conju
Recalled July 6, 2016 · FDA recall Z-2084-2016 · Integra LifeSciences
Hazard
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter. · FDA Class II
Units
182 units
Sold at
Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA
Description
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.
Product
K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws
Manufacturer(s)
Integra LifeSciences Corp.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →