RecallWatchUSA

Home / Recalls / Medical Devices

Integra York PA, Inc recalls Rochester-Pean Hemostatic Forceps 61/4, Curved, Sterile. The Rochester-Pean Hemostatic Forceps 61/4 Curved are single

Recalled December 21, 2016 · FDA recall Z-0751-2017 · Integra York PA

Hazard
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument. · FDA Class II
Units
150 units
Sold at
US Distribution in NC and TX.

Description

The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.

Product
Rochester-Pean Hemostatic Forceps 61/4, Curved, Sterile. The Rochester-Pean Hemostatic Forceps 61/4 Curved are single-use sterile surgical instruments intended for cutting and grasping tissue in general surgical procedures.
Manufacturer(s)
Integra York PA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Integra York PA, Inc recalls Rochester-Pean Hemostatic Forceps 61/4, Curved, Sterile. The Rochester-Pean Hemostatic Forceps 61/4 Curved are single | RecallWatchUSA