Home / Recalls / Medical Devices
Integra York PA, Inc recalls Rochester-Pean Hemostatic Forceps 61/4, Curved, Sterile. The Rochester-Pean Hemostatic Forceps 61/4 Curved are single
Recalled December 21, 2016 · FDA recall Z-0751-2017 · Integra York PA
Hazard
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument. · FDA Class II
Units
150 units
Sold at
US Distribution in NC and TX.
Description
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
Product
Rochester-Pean Hemostatic Forceps 61/4, Curved, Sterile. The Rochester-Pean Hemostatic Forceps 61/4 Curved are single-use sterile surgical instruments intended for cutting and grasping tissue in general surgical procedures.
Manufacturer(s)
Integra York PA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →