RecallWatchUSA

Home / Recalls / Medical Devices

Intelerad Medical Systems recalls IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and

Recalled August 27, 2014 · FDA recall Z-2267-2014 · Intelerad Medical Systems

Hazard
In specific situation a partially truncated file may be written to the IntelePACS server. This partially truncated file still can be viewed by InteleViewer and InteleViewer will report to the user that the file in question is complete. This situation only occurs if a file is partially truncated. · FDA Class II
Units
230
Sold at
Worldwide Distribution - USA Nationwide and the countries of Scotland, Ireland, United Kingdom, Canada, Australia, and New Zealand.

Description

In specific situation a partially truncated file may be written to the IntelePACS server. This partially truncated file still can be viewed by InteleViewer and InteleViewer will report to the user that the file in question is complete. This situation only occurs if a file is partially truncated.

Product
IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.
Manufacturer(s)
Intelerad Medical Systems
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Intelerad Medical Systems recalls IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and | RecallWatchUSA