International Biophysics Corp. recalls IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loo

Recalled August 26, 2015 · FDA recall Z-2424-2015 · International Biophysics

Hazard
It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria. · FDA Class II
Units
771 units
Sold at
Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa

Description

It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.

Product
IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loop is a single-use, disposable loop used in occluding, retracting, and identifying arteries, veins, tendons, and nerves in surgical procedures.
Manufacturer(s)
International Biophysics Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

International Biophysics Corp. recalls IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loo | RecallWatchUSA