Intuitive Surgical, Inc. recalls da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
Recalled June 3, 2020 · FDA recall Z-2131-2020 · Intuitive Surgical
Hazard
Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form. · FDA Class II
Units
305 units
Sold at
Distribution US nationwide and South Korea.
Description
Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
Product
da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
Manufacturer(s)
Intuitive Surgical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →