RecallWatchUSA

Home / Recalls / Medical Devices

Intuitive Surgical, Inc. recalls da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11

Recalled March 9, 2022 · FDA recall Z-0702-2022 · Intuitive Surgical

Hazard
Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain. · FDA Class II
Units
149,646 reducers
Sold at
Worldwide distribution - US Nationwide distribution in the states of California, Wisconsin, Virginia, Ohio, Texas, New York, Nevada, Connecticut, New Jersey, Massachusetts, Georgia, Florida, Indiana, North Carolina, Missouri, Pennsylvania, Michigan, Oregon, Kentucky, Hawaii, Tennessee, Alabama, Nebraska, Maryland, Oklahoma, Colorado, Arkansas, Arizona, Kansas, Illinois, South Carolina, Minnesota, North Dakota, Iowa, Louisiana, Washington, Delaware, Mississippi, New Mexico, District of Columbia, Idaho, New Hampshire, Rhode Island, West Virginia, South Dakota, Alaska, Maine, Utah, Montana, Vermont, Wyoming, Puerto Rico and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Luxembourg, Martinique, Monaco, Netherlands, Norway, Poland, Qatar, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

Description

Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.

Product
da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11
Manufacturer(s)
Intuitive Surgical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Intuitive Surgical, Inc. recalls da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11 | RecallWatchUSA