RecallWatchUSA

Home / Recalls / Medical Devices

Intuitive Surgical, Inc. recalls The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.

Recalled August 22, 2018 · FDA recall Z-2835-2018 · Intuitive Surgical

Hazard
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope. · FDA Class II
Units
1
Sold at
U.S.

Description

An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.

Product
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Manufacturer(s)
Intuitive Surgical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Intuitive Surgical, Inc. recalls The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16. | RecallWatchUSA