RecallWatchUSA

Home / Recalls / Medical Devices

Invacare Corporation recalls Aquatec Ocean VIP, Aquatec Ocean Dual-VIP, Aquatec Ocean E-VIP mobile shower and toilet commode with tilting function.

Recalled November 13, 2013 · FDA recall Z-0167-2014 · Invacare

Hazard
The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Ocean Dual VIP do not withstand a required pull out force of 500 N, when the product is used in a reclined position and or tilted position. This may result in the snap-on fixture coming unlatched. · FDA Class II
Units
538 (US only) 3317 (Total)
Sold at
Worldwide Distribution - USA including PA and Internationally to Canada.

Description

The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Ocean Dual VIP do not withstand a required pull out force of 500 N, when the product is used in a reclined position and or tilted position. This may result in the snap-on fixture coming unlatched.

Product
Aquatec Ocean VIP, Aquatec Ocean Dual-VIP, Aquatec Ocean E-VIP mobile shower and toilet commode with tilting function. Mobile shower and toilet commode with tilting function
Manufacturer(s)
Invacare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov