Home / Recalls / Medical Devices
Invacare Corporation recalls DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1
Recalled August 30, 2023 · FDA recall Z-2445-2023 · Invacare
Hazard
Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user · FDA Class II
Units
586 devices
Sold at
US Nationwide distribution.
Description
Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user
Product
DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE GLOSS 680 ROLLATOR 1655081 - DOLOMITE GLOSS 600 ROLLATOR 1655082 - DOLOMITE GLOSS 520 ROLLATOR 1655083 - DOLOMITE GLOSS 450 ROLLATOR
Manufacturer(s)
Invacare Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →