RecallWatchUSA

Home / Recalls / Medical Devices

Invacare Corporation recalls Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patie

Recalled December 12, 2018 · FDA recall Z-0549-2019 · Invacare

Hazard
Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained. Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur. For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures. · FDA Class II
Units
73,703 units
Sold at
Worldwide Distribution: US (Nationwide) and countries of: Canada, Colombia, Germany, Guam, India, Northern Mariana Islands, Philippines and South Africa.

Description

Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained. Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur. For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures.

Product
Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patient lift is to lift and transfer a patient from one surface to another, as from a bed to a wheelchair. The device includes straps and a sling to support the patient.
Manufacturer(s)
Invacare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Invacare Corporation recalls Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patie | RecallWatchUSA