Iradimed Corporation recalls MRidium 1145 Dose Reduction System (DERS) drug library kit. (This is an option for the MRidium 3860+MR Infusion Pump Sy
Recalled August 14, 2013 · FDA recall Z-1874-2013 · Iradimed
Hazard
The Dose Error Reduction System (DERS) can indicate an incorrect recommended value for the rate via a specific key sequence during the initial infusion setup. This can result in a risk of over-infusion or under-infusion if the infusion is started with this rate value. · FDA Class I
Units
176 individuals kits
Sold at
Worldwide distribution: US (nationwide) including states of: AL, CA, CO, FL, GA, IL, KY, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, RI, TN, TX, UT, VA, and WA; and countries of: Australia, Austria, Denmark, Epsom, France, Germany, London, Netherlands, Saudia Arabia, and United Kingdom.
Description
The Dose Error Reduction System (DERS) can indicate an incorrect recommended value for the rate via a specific key sequence during the initial infusion setup. This can result in a risk of over-infusion or under-infusion if the infusion is started with this rate value.
Product
MRidium 1145 Dose Reduction System (DERS) drug library kit. (This is an option for the MRidium 3860+MR Infusion Pump System). The MRidium 3860+MRI Infusion pump are intended for general hospital or clinical use by medical professional whenever it is required to infuse patients with intra-venous or intra-arterial fluids before, during or after magnetic Resonance imagining (MRI) scans.
Manufacturer(s)
Iradimed Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →