RecallWatchUSA

Home / Recalls / Medical Devices

IsoTis OrthoBiologics, Inc. recalls OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095

Recalled August 5, 2026 · FDA recall Z-2820-2026 · IsoTis OrthoBiologics

Hazard
Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth. · FDA Class II
Units
98
Sold at
US Nationwide distribution in the states of AK, NJ, TN, AR, CA, WA, OH.

Description

Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.

Product
OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
Manufacturer(s)
IsoTis OrthoBiologics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov