Home / Recalls / Medical Devices
Joerns Healthcare recalls ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (
Recalled June 6, 2018 · FDA recall Z-2034-2018 · Joerns Healthcare
Hazard
Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells. · FDA Class II
Units
46 units
Sold at
Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland
Description
Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.
Product
ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).
Manufacturer(s)
Joerns Healthcare
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →