Home / Recalls / Medical Devices
Karl Storz Endoscopy recalls Flexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: are intended to
Recalled September 2, 2020 · FDA recall Z-2895-2020 · Karl Storz Endoscopy
Hazard
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection. · FDA Class II
Units
428
Sold at
Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.
Description
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
Product
Flexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Manufacturer(s)
Karl Storz Endoscopy
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →