Home / Recalls / Medical Devices
KCI USA, Inc. recalls BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or d
Recalled October 3, 2012 · FDA recall Z-2466-2012 · KCI USA
Hazard
KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow · FDA Class II
Units
575
Sold at
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
Description
KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow
Product
BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy.
Manufacturer(s)
KCI USA, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →