Kentec Medical, Inc recalls Ameritus Medical Products ENFit Polyurethane and Silicone Feeding Tubes. Sterile and single use only. Catalog number ENF
Recalled July 15, 2020 · FDA recall Z-2514-2020 · Kentec Medical
Hazard
Feeding tube plastic tethered closure cap may possibly fracture or break. Possible health hazards: (1) Enteral feeding and medication administration could be possibly delayed, until replacement tube is placed into patient (2) Patient could be unnecessarily overexposed to radiation if multiple x-rays are needed to confirm placement of a new tube, if x-ray is used as the placement confirmation method. · FDA Class II
Units
480
Sold at
US: CA, IA, and VA
Description
Feeding tube plastic tethered closure cap may possibly fracture or break. Possible health hazards: (1) Enteral feeding and medication administration could be possibly delayed, until replacement tube is placed into patient (2) Patient could be unnecessarily overexposed to radiation if multiple x-rays are needed to confirm placement of a new tube, if x-ray is used as the placement confirmation method.
Product
Ameritus Medical Products ENFit Polyurethane and Silicone Feeding Tubes. Sterile and single use only. Catalog number ENF-40P-50 (4 fr. 50 cm ENFit Enteral Feeding Tube).
Manufacturer(s)
Kentec Medical, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →