Home / Recalls / Medical Devices
Keystone Dental Inc recalls Paltop 1.25 Hex Drivers, Short Part Number: 60-70102
Recalled April 3, 2019 · FDA recall Z-1064-2019 · Keystone Dental
Hazard
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature · FDA Class II
Units
3 Hex Driver
Sold at
FL, MA, SC, VA
Description
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Product
Paltop 1.25 Hex Drivers, Short Part Number: 60-70102
Manufacturer(s)
Keystone Dental Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →