RecallWatchUSA

Home / Recalls / Medical Devices

Ki Mobility Llc recalls Ki Mobility, Little Wave Tyke (Weight Limit: 95 lbs), Specifically the IFU The Little Wave Tyke is a manually operated

Recalled January 19, 2022 · FDA recall Z-0467-2022 · Ki Mobility

Hazard
The Instructions for Use provided with the device contained information intended for use of a mechanical wheelchair, not a chair with casters. · FDA Class III
Units
33 units shipped with affected manuals
Sold at
Domestic: IL, IN, KY, MI, MO, NC, SC, and TN.

Description

The Instructions for Use provided with the device contained information intended for use of a mechanical wheelchair, not a chair with casters.

Product
Ki Mobility, Little Wave Tyke (Weight Limit: 95 lbs), Specifically the IFU The Little Wave Tyke is a manually operated device that can be used for medical purposes and is used to assist a disabled person to move from one place to another while sitting. Intended to be used by persons up to 95 pounds for transport.
Manufacturer(s)
Ki Mobility Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov