RecallWatchUSA

Home / Recalls / Medical Devices

Kimberly-Clark Corporation recalls Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000

Recalled August 15, 2012 · FDA recall Z-2181-2012 · Kimberly-Clark

Hazard
A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke. · FDA Class II
Units
129 units
Sold at
Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.

Description

A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.

Product
Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Manufacturer(s)
Kimberly-Clark Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Kimberly-Clark Corporation recalls Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 | RecallWatchUSA