Konica Minolta Medical Imaging USA, Inc. recalls ImagePilot, Model No. D9MA

Recalled July 4, 2018 · FDA recall Z-2282-2018 · Konica Minolta Medical Imaging USA

Hazard
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution. · FDA Class II
Sold at
Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.

Description

Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

Product
ImagePilot, Model No. D9MA
Manufacturer(s)
Konica Minolta Medical Imaging USA, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov