Laboratoires Bodycad, Inc. recalls Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant a
Recalled November 24, 2021 · FDA recall Z-0244-2022 · Laboratoires Bodycad
Hazard
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use. · FDA Class II
Units
10 in the U.S.
Sold at
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Description
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Product
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Manufacturer(s)
Laboratoires Bodycad, Inc.
Made in
Canada