RecallWatchUSA

Home / Recalls / Medical Devices

Landauer recalls Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SE

Recalled September 20, 2023 · FDA recall Z-2557-2023 · Landauer

Hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. · FDA Class II
Units
5958 units
Sold at
Worldwide

Description

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Product
Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR readers
Manufacturer(s)
Landauer

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Landauer recalls Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SE | RecallWatchUSA