Home / Recalls / Medical Devices
Landauer recalls Constancy (Cs-137) nanoDot D2DNN, Model Numbers: a) 18155-000 (EMEA number VKITCON001); b) 18155-KIT; radiation mon
Recalled September 20, 2023 · FDA recall Z-2572-2023 · Landauer
Hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. · FDA Class II
Units
1880 units
Sold at
Worldwide
Description
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Product
Constancy (Cs-137) nanoDot D2DNN, Model Numbers: a) 18155-000 (EMEA number VKITCON001); b) 18155-KIT; radiation monitoring dosimeter used with the microSTAR readers
Manufacturer(s)
Landauer
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →