Landauer recalls nanoDot D2DXS, Model Numbers: a) 03062-1MO; b) 03062-3MO; c) 03062-OTO; d) 03062-SMO; radiation monitoring dosi

Recalled September 20, 2023 · FDA recall Z-2547-2023 · Landauer

Hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. · FDA Class II
Units
0
Sold at
Worldwide

Description

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Product
nanoDot D2DXS, Model Numbers: a) 03062-1MO; b) 03062-3MO; c) 03062-OTO; d) 03062-SMO; radiation monitoring dosimeter used with the microSTAR readers
Manufacturer(s)
Landauer