RecallWatchUSA

Home / Recalls / Medical Devices

Landauer recalls nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR r

Recalled September 20, 2023 · FDA recall Z-2553-2023 · Landauer

Hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. · FDA Class II
Units
16760 units
Sold at
Worldwide

Description

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Product
nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
Manufacturer(s)
Landauer