Leica Microsystems, Inc. recalls Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microsco

Recalled October 10, 2018 · FDA recall Z-0019-2019 · Leica Microsystems

Hazard
Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock. · FDA Class II
Units
1 unit
Sold at
Distribution in US state of MO.

Description

Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.

Product
Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.
Manufacturer(s)
Leica Microsystems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov