Home / Recalls / Medical Devices
Leica Microsystems, Inc. recalls Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
Recalled December 21, 2022 · FDA recall Z-0476-2023 · Leica Microsystems
Hazard
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown. · FDA Class II
Units
38 units
Sold at
US Nationwide
Description
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
Product
Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
Manufacturer(s)
Leica Microsystems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →