RecallWatchUSA

Home / Recalls / Medical Devices

Leica Microsystems, Inc. recalls Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063

Recalled October 5, 2022 · FDA recall Z-1848-2022 · Leica Microsystems

Hazard
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used. · FDA Class II
Units
30 devices
Sold at
US Nationwide distribution.

Description

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Product
Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063
Manufacturer(s)
Leica Microsystems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov