Leica Microsystems, Inc. recalls Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supe

Recalled June 11, 2014 · FDA recall Z-1698-2014 · Leica Microsystems

Hazard
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity. · FDA Class II
Units
33 units
Sold at
Nationwide Distribution including AZ, CA, FL, and NY.

Description

Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.

Product
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Manufacturer(s)
Leica Microsystems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov