Life Technologies Holdings Pte Ltd recalls Torrent Suite Dx Software Version 5.8 Catalog Number:A36601 and A36602 OEM Customer software: Torrent Suite Dx T430 S
Recalled October 30, 2019 · FDA recall Z-0188-2020 · Life Technologies Holdings Pte
Hazard
Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant · FDA Class II
Units
total of Ion+vela +chassis =400 units
Sold at
states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland
Description
Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant
Product
Torrent Suite Dx Software Version 5.8 Catalog Number:A36601 and A36602 OEM Customer software: Torrent Suite Dx T430 Software v5.0 DARUI DR-SEQ 800 Software UDI:(01)10190302014090 (11)000000(10)5.8 (240)A36601
Manufacturer(s)
Life Technologies Holdings Pte Ltd
Made in
Singapore