Home / Recalls / Medical Devices
Lin-Zhi International Inc recalls Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent
Recalled May 8, 2019 · FDA recall Z-1256-2019 · Lin-Zhi International
Hazard
Shelf life of the product may be reduced due to degradation of the assay. · FDA Class III
Units
49 kits
Sold at
US Nationwide Distribution in the states of CA, MA, GA, KY, NC, MI, MD, NC, CT, AZ SC, TN, OR
Description
Shelf life of the product may be reduced due to degradation of the assay.
Product
Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent causes faster degradation in EtOH assay. Possible container or pitchers with some detergent residue was used while producing R2 of the defective lot.
Manufacturer(s)
Lin-Zhi International Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →