RecallWatchUSA

Home / Recalls / Medical Devices

Lin-Zhi International Inc recalls Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin

Recalled April 2, 2014 · FDA recall Z-1270-2014 · Lin-Zhi International

Hazard
Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL). · FDA Class III
Units
120 units, 60 of each lot number
Sold at
US Distribution including CA.

Description

Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

Product
Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer
Manufacturer(s)
Lin-Zhi International Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Lin-Zhi International Inc recalls Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin | RecallWatchUSA