RecallWatchUSA

Home / Recalls / Medical Devices

LINK BIO CORP recalls LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

Recalled July 31, 2024 · FDA recall Z-2445-2024 · LINK BIO

Hazard
The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process. · FDA Class II
Units
8 units
Sold at
US Nationwide distribution in the states of FL, GA, KS, OH, WI.

Description

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.

Product
LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00
Manufacturer(s)
LINK BIO CORP