Home / Recalls / Medical Devices
LITTLE RAPIDS CORPORATION recalls EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3
Recalled February 21, 2018 · FDA recall Z-0578-2018 · LITTLE RAPIDS
Hazard
The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through. · FDA Class II
Units
153 cases, 30 units in each case
Sold at
TX, CT, TN, MO, WI, FL, SD, MS, CA, MS
Description
The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.
Product
EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3
Manufacturer(s)
LITTLE RAPIDS CORPORATION
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →