Home / Recalls / Medical Devices
Luminex Corporation recalls Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Recalled May 25, 2022 · FDA recall Z-1092-2022 · Luminex
Hazard
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results. · FDA Class II
Units
24 units
Sold at
US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None
Description
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
Product
Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Manufacturer(s)
Luminex Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →