RecallWatchUSA

Home / Recalls / Medical Devices

Magellan Diagnostics, Inc. recalls LeadCare Plus Blood Lead Analyzer Part Number: 82-0001 Intended for the quantitative measurement of lead in a whole

Recalled April 11, 2018 · FDA recall Z-1296-2018 · Magellan Diagnostics

Hazard
The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot. · FDA Class II
Units
79
Sold at
Worldwide Distribution

Description

The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.

Product
LeadCare Plus Blood Lead Analyzer Part Number: 82-0001 Intended for the quantitative measurement of lead in a whole blood sample.
Manufacturer(s)
Magellan Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov