Home / Recalls / Medical Devices
Magellan Diagnostics, Inc. recalls LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following mat
Recalled April 29, 2026 · FDA recall Z-1894-2026 · Magellan Diagnostics
Hazard
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing. · FDA Class II
Units
136
Sold at
Worldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, South Korea, Spain, UAE, Uruguay, Vietnam.
Description
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Product
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Manufacturer(s)
Magellan Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →