RecallWatchUSA

Home / Recalls / Medical Devices

Magellan Diagnostics, Inc. recalls LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098

Recalled July 7, 2021 · FDA recall Z-1954-2021 · Magellan Diagnostics

Hazard
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples. · FDA Class I
Units
441 kits (192 tests/kit = 84,672 tests)
Sold at
Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.

Description

Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.

Product
LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098
Manufacturer(s)
Magellan Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov