Home / Recalls / Medical Devices
Maquet Cardiovascular, LLC recalls Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 Adult ECC Pack (2)Material: 701065765R02 BO-TOP 36504 Adult ECC
Recalled June 5, 2019 · FDA recall Z-1663-2019 · Maquet Cardiovascular
Hazard
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured · FDA Class II
Units
N/A
Sold at
Nationwide
Description
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Product
Maquet Getinge- (1)Material: 701065765R01 BO-TOP 36504 Adult ECC Pack (2)Material: 701065765R02 BO-TOP 36504 Adult ECC Pack
Manufacturer(s)
Maquet Cardiovascular, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →