Home / Recalls / Medical Devices
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 15800 TUBING PACK Material: 3000065087
Recalled June 5, 2019 · FDA recall Z-1498-2019 · Maquet Cardiovascular
Hazard
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured · FDA Class II
Units
N/A
Sold at
Nationwide
Description
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Product
Maquet Getinge-BEQ-TOP 15800 TUBING PACK Material: 3000065087
Manufacturer(s)
Maquet Cardiovascular, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →