RecallWatchUSA

Home / Recalls / Medical Devices

Maquet Cardiovascular, LLC recalls Maquet Getinge-BO-T 51502 Ochsner ECC Adult Pack Material: 709000385

Recalled June 5, 2019 · FDA recall Z-1614-2019 · Maquet Cardiovascular

Hazard
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured · FDA Class II
Units
N/A
Sold at
Nationwide

Description

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Product
Maquet Getinge-BO-T 51502 Ochsner ECC Adult Pack Material: 709000385
Manufacturer(s)
Maquet Cardiovascular, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Maquet Cardiovascular, LLC recalls Maquet Getinge-BO-T 51502 Ochsner ECC Adult Pack Material: 709000385 | RecallWatchUSA