Home / Recalls / Medical Devices
Maquet Cardiovascular, LLC recalls Vasoshield Syringe Packs, Model VH-5001
Recalled January 25, 2023 · FDA recall Z-0996-2023 · Maquet Cardiovascular
Hazard
Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event. · FDA Class II
Units
530 units US, 215 units OUS
Sold at
US Nationwide. Japan, Germany.
Description
Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.
Product
Vasoshield Syringe Packs, Model VH-5001
Manufacturer(s)
Maquet Cardiovascular, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →