RecallWatchUSA

Home / Recalls / Medical Devices

Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel H

Recalled October 1, 2025 · FDA recall Z-2636-2025 · Maquet Cardiovascular

Hazard
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws. · FDA Class II
Units
98 units
Sold at
US Nationwide distribution in the states of FL, IL, MD & TX.

Description

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

Product
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
Manufacturer(s)
Maquet Cardiovascular, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel H | RecallWatchUSA