Maquet Cardiovascular Us Sales, Llc recalls FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ven
Recalled July 18, 2012 · FDA recall Z-1925-2012 · Maquet Cardiovascular Us Sales
Hazard
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position. · FDA Class I
Units
284 units
Sold at
Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Description
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
Product
FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
Manufacturer(s)
Maquet Cardiovascular Us Sales, Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →